Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing Recalled by Siemens Medical Solutions USA, Inc Due to Firm has become aware of an unintended behavior...

Date: June 20, 2012
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System

Quantity: 15

Why Was This Recalled?

Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could