Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now Recalled by Siemens Medical Solutions USA, Inc Due to Software error may result in sporadic problems causing...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.
Affected Products
SIEMENS SOMATOM with software syngo CT VA30A_SP4 Models: SOMATOM go.Now 11061610 & 11061618 SOMATOM go.Up 11061620 & 11061628 SOMATOM go.All 11061630 & 11061638 SOMATOM go.Top 11061640 & 11061648 SOMATOM go.Sim 11061660 & 11061668 SOMATOM go.Open Pro 11061670 & 11061678 SOMATOM X.cite 11330001 Intended Use: Computed tomography x-ray system
Quantity: 547 units US
Why Was This Recalled?
Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Medical Solutions USA, Inc
Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report