Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens SimView NT and SimView NT with IDI simulation system Recalled by Siemens Medical Solutions USA, Inc Due to Siemens became aware of a very rare occurrence...

Date: October 10, 2014
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Siemens SimView NT and SimView NT with IDI simulation system components: Product Usage: The intended use of the Radiation Therapy Simulation system is used for simulation of treatment planning prior to the therapeutic treatment of cancer.

Quantity: 6

Why Was This Recalled?

Siemens became aware of a very rare occurrence where an intended movement of the SimView NT simulation system components could not be stopped with usual hand control commands or by use of the collision detection functionality. It is possible that simulation system components could collide with patient and/or system equipment.