Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens ACUSON X700 Ultrasound Systems with software version 1.0.00 Recalled by Siemens Medical Solutions USA, Inc. Due to When fourSight 4D ultrasound imaging, 3-Scrape real time...

Date: April 2, 2014
Company: Siemens Medical Solutions USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc. directly.

Affected Products

Siemens ACUSON X700 Ultrasound Systems with software version 1.0.00, 1.0.01, 1.0.02, 1.1.00 and 1.1.02. ultrasound imaging system.

Quantity: 490

Why Was This Recalled?

When fourSight 4D ultrasound imaging, 3-Scrape real time 3D imaging, or Anatomical M-mode are used during an Obstetric or Gynecology study, data from the previous patient could replace the current patient's date in the report.

Where Was This Sold?

This product was distributed nationwide across all 50 states.