Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MAMMOMAT Revelation Recalled by Siemens Medical Solutions USA, Inc Due to If an examination is interrupted when the InSpect...

Date: May 7, 2020
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

MAMMOMAT Revelation, model no. 11343300 - Product Usage: is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals.

Quantity: 230 worldwide (64 US)

Why Was This Recalled?

If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report