Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Interventional Fluoroscopic X-Ray System - Product Usage: developed for single Recalled by Siemens Medical Solutions USA, Inc Due to Siemens has become aware of a potential issue...

Date: February 26, 2021
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Interventional Fluoroscopic X-Ray System - Product Usage: developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Quantity: 29 systems

Why Was This Recalled?

Siemens has become aware of a potential issue in which the imaging system UPS sporadically forces a shutdown of the imaging system PC without a true power supply problem.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.