Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

COHERENCE Oncologist Recalled by Siemens Medical Solutions USA, Inc Due to A potential safety issue arises when an offset...

Date: January 16, 2020
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

COHERENCE Oncologist, Model Nos. 07333680 & 07351898

Quantity: 64 Worldwide/26 Domestic

Why Was This Recalled?

A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.

Where Was This Sold?

This product was distributed to 16 states: CA, FL, GA, IA, MA, MI, NE, NV, NJ, NY, OH, OK, PA, SD, TN, WV

Affected (16 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report