Cios Alpha with software version VA30-mobile C-arm x-ray Material # Recalled by Siemens Medical Solutions USA, Inc Due to Continuous operation of the Cios Alpha at high...

Date: March 22, 2019
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients

Quantity: 2 US

Why Was This Recalled?

Continuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report