Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

AXIOM Luminos dRF system with SW VD10 or on Luminos Recalled by Siemens Medical Solutions USA, Inc Due to It was determined that a potential malfunction may...

Date: August 12, 2014
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

AXIOM Luminos dRF system with SW VD10 or on Luminos dRF Max The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Quantity: 33

Why Was This Recalled?

It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel and result in malfunction. In the very unlikely case of fluid infiltration a permanent short circuit within the control panel of the Luminos dRF / Luminos dRF Max will automatically d

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions USA, Inc

Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report