Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARTISTE with syngo RT Therapist version 4.3 Product Usage: The Recalled by Siemens Medical Solutions USA, Inc Due to The potential for a safety issue, for the...

Date: May 29, 2013
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

ARTISTE with syngo RT Therapist version 4.3 Product Usage: The intended use of the SIEMENS branded ARTISTE family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

Quantity: 8 units

Why Was This Recalled?

The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure MV or In-Line kView images using the syngo RT Therapist 4.3.

Where Was This Sold?

Worldwide distribution