Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283) Recalled by Siemens Medical Solutions USA, Inc Due to It may occur that after system startup no...

Date: May 24, 2022
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System

Quantity: 233 worldwide; 32 US

Why Was This Recalled?

It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

Where Was This Sold?

This product was distributed nationwide across all 50 states.