Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Artis zee systems x-ray Recalled by Siemens Medical Solutions USA, Inc Due to The firm became aware of a potential issue...

Date: January 21, 2013
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Artis zee systems x-ray, angiographic system

Quantity: 17

Why Was This Recalled?

The firm became aware of a potential issue with Artis zee systems bearing model numbers 10094135, 10094137, and 10094141 equipped with a card collimator from a specific lot where a dislocation of the scatter radiation filter inside the card collimator may occur.

Where Was This Sold?

This product was distributed to 14 states: AZ, AR, CA, CT, FL, IA, KS, KY, MO, MT, NY, OK, PA, WV

Affected (14 states)Not affected