Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Artis Zee and Zeego Recalled by Siemens Medical Solutions USA, Inc Due to This recall has been initiated due to a...

Date: October 26, 2018
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

Artis Zee and Zeego, model no. 10272462 AXIOM Artis zee/Zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Quantity: 64 (16 in US)

Why Was This Recalled?

This recall has been initiated due to a possible leakage of a hose in the cooling unit of the X-ray tube. The Artis zee system may lose coolant due to a possible leakage in a hose of the cooling unit in the X-ray tube. This will cause the X-ray tube not to be sufficiently cooled down. As a result, the system will display a message Tube hot, have a break . The X-ray will be blocked several minutes later and to prevent any damage to the machine the following message will be displayed NO XRAY, TUBE TOO HOT . This issue is very sporadic and may occur during an ongoing clinical procedure. Use of this product may have a potential impact on planned procedures as they may be delayed as well as in emergent cases, as clinical treatment may need to be terminated and transferred to a functioning system.

Where Was This Sold?

This product was distributed to 12 states: AR, CA, KS, KY, MD, MN, NM, OK, PA, RI, TX, UT

Affected (12 states)Not affected