Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARTIS pheno systems with a Siemens Healthineers table or a Recalled by Siemens Medical Solutions USA, Inc Due to In the event of any unintended table movement,...

Date: February 2, 2023
Company: Siemens Medical Solutions USA, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

ARTIS pheno systems with a Siemens Healthineers table or a Trumpf/ MAQUET table-A Fluoroscopic X-Ray system. Material # 10849000.

Quantity: 141 units US

Why Was This Recalled?

In the event of any unintended table movement, the system may not detect the incorrect direction, could lead to the injury of a patient, staff member, operator, or equipment.

Where Was This Sold?

This product was distributed nationwide across all 50 states.