Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET Recalled by Siemens Medical Solutions USA, Inc Due to If, during the procedure, X-ray has been released...

Date: February 2, 2023
Company: Siemens Medical Solutions USA, Inc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

ARTIS pheno systems with a Siemens Healthineers table or Trumpf/MAQUET table (OEM)- Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Model: 10849000

Quantity: 141 units US

Why Was This Recalled?

If, during the procedure, X-ray has been released and a reference image has been stored, the following issue may occur: If "Adjust C-arm to Ref" is activated when the C-Arm is positioned outside of the working range the C-Arm will reach the target position with an inaccuracy of 5-10 mm. As a result, the message Endposition reached will be displayed .Leading to unintended direction of the movement causing crushing of a patient, staff member, operator, or equipmen Live images may not match the previously stored reference images. Overlay images may be shown inaccurate on anatomy (e.g., DSA Roadmap workflow does not match real anatomy). This may cause e.g., a vessel perforation in DSA roadmap.

Where Was This Sold?

This product was distributed nationwide across all 50 states.