Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Artis icono floor in combination with a Siemens Healthineers table Recalled by Siemens Medical Solutions USA, Inc Due to Five potential software issues affecting Artis pheno and...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.
Affected Products
Artis icono floor in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). Intended use single and biplane diagnostic imaging and interventional procedures. Model: 11327700
Quantity: 35 units
Why Was This Recalled?
Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Medical Solutions USA, Inc
Siemens Medical Solutions USA, Inc has 613 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report