Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ARTIS Icono biplane Model:11327600 ARTIS Icono floor system Model: 11327700 Recalled by Siemens Medical Solutions USA, Inc Due to Misleading error messages and a gap in the...

Date: December 18, 2020
Company: Siemens Medical Solutions USA, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc directly.

Affected Products

ARTIS Icono biplane Model:11327600 ARTIS Icono floor system Model: 11327700 with software version VE20B - Product Usage: Intended for single and biplane diagnostic imaging and interventional procedures.

Quantity: 41 U.S

Why Was This Recalled?

Misleading error messages and a gap in the Operator Manual which affects ARTIS Icono biplane or ARTIS Icono floor systems with software version VE20B. Potential issues include System error management, Erroneous error messages, Zoom/Pan Function, Grid Indication, and Coolant Level. May cause procedures to be terminated and performed on an alternative x-ray system