Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ACUSON SC2000 Ultrasound System Product Usage: The SC2000 ultrasound imaging Recalled by Siemens Medical Solutions USA, Inc. Due to The application may underestimate the EROA (Effective Regurgitant...

Date: January 26, 2018
Company: Siemens Medical Solutions USA, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc. directly.

Affected Products

ACUSON SC2000 Ultrasound System Product Usage: The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Quantity: 564 systems

Why Was This Recalled?

The application may underestimate the EROA (Effective Regurgitant Orifice Area) in comparison to the same patient results obtained with the 4Z1c volume transthoracic echocardiography transducer.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Medical Solutions USA, Inc.

Siemens Medical Solutions USA, Inc. has 87 total recalls tracked by RecallDetector.