Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ACUSON Redwood 2.0 ultrasound systems Recalled by Siemens Medical Solutions USA, Inc. Due to If a user-generated preset for an 18L6 transducer...

Date: July 13, 2023
Company: Siemens Medical Solutions USA, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Medical Solutions USA, Inc. directly.

Affected Products

ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

Quantity: 796

Why Was This Recalled?

If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

Where Was This Sold?

This product was distributed nationwide across all 50 states.