Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Syva Emit Methotrexate Assay Recalled by Siemens Healthcare Diagnostics, Inc. Due to Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate...

Date: March 8, 2016
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Syva Emit Methotrexate Assay, is a homogeneous enzyme immunoassay intended for use in the quantitative analysis of methotrexate in human serum or plasma.

Quantity: 18364 units

Why Was This Recalled?

Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that values of methotrexate for patient and quality control samples increased.

Where Was This Sold?

This product was distributed to 12 states: CA, FL, GA, MI, MT, NJ, NY, NC, PA, TX, VA, DC

Affected (12 states)Not affected