Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Recalled by Siemens Healthcare Diagnostics, Inc. Due to Under certain conditions the system may not perform...

Date: May 10, 2013
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.

Quantity: 262

Why Was This Recalled?

Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.