Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens IMMULITE 2000 Immunoassay System - In vitro Chemistry Analyzer Recalled by Siemens Healthcare Diagnostics Inc. Due to Flexible tubing connected to the liquid waste bottle...

Date: May 15, 2020
Company: Siemens Healthcare Diagnostics Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics Inc. directly.

Affected Products

Siemens IMMULITE 2000 Immunoassay System - In vitro Chemistry Analyzer SMN #10380062 - Product Usage: are intended for professional use in a laboratory environment only.

Quantity: 1555 units

Why Was This Recalled?

Flexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard

Where Was This Sold?

This product was distributed nationwide across all 50 states.