Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens Healthcare ADVIA Centaur Tnl-Ultra Assay . In-Vitro Diagnostic for Recalled by Siemens Healthcare Diagnostics, Inc Due to Exhibits a greater than 10% change in results...

Date: May 2, 2016
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

Siemens Healthcare ADVIA Centaur Tnl-Ultra Assay . In-Vitro Diagnostic for quantitative determination of cardiac troponin For in vitro diagnostic use in the quantitative determination of cardiac troponin I (cTnl) in serum, heparinized plasma, or EDTA plasma using the ADVIA Centaur/ADVIA Centaur XP/ADVIA Centaur XPT/ADVIA Centaur CP systems Cat No. 02789602/SMN 1031 7708 (100 test) Cat No. 02790309/SMN 10317709 (500 test)

Quantity: SMN 10317708 = 16434 kits and SMN 10317709 = 16033 kits

Why Was This Recalled?

Exhibits a greater than 10% change in results in samples with biotin levels up to 10 ng/mL (41 nmol/L). Instructions for Use (IFU) states that specimens that have up to 10 ng/mL (41 nmol/L) of biotin

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected