Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain Recalled by Siemens Healthcare Diagnostics, Inc. Due to A manufacturing defect causes Atellica CH Reaction Ring...

Date: October 21, 2024
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction volumes and present them to each processing step. Siemens Material Number(SMN): 11099326 All lots starting with N15 or 15 are expected to have the defect.

Quantity: 9729 units

Why Was This Recalled?

A manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected