Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 Recalled by Siemens Healthcare Diagnostics, Inc Due to Serum controls may go out of range low...

Date: March 25, 2014
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat No. 06367974 SMN 10310308 (100 tests) Cat No. 00203473 SMN 10331250 (Ref) (500 tests) Cat No. 09132781 SMN 10340209 (Ref) (2500 tests) For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems

Quantity: 49123 units

Why Was This Recalled?

Serum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc

Siemens Healthcare Diagnostics, Inc has 118 total recalls tracked by RecallDetector.