Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number Recalled by Siemens Healthcare Diagnostics, Inc Due to Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit...

Date: December 7, 2018
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number (SMN): 10492398

Quantity: 10673 kits

Why Was This Recalled?

Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected