Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

QuikLYTE Dilution Check: used to monitor and adjust for the Recalled by Siemens Healthcare Diagnostics, Inc. Due to A positive bias beyond 5% resulting in a...

Date: February 24, 2015
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

QuikLYTE Dilution Check: used to monitor and adjust for the dilution ratio on Dimension clinical chemistry systems with the QuikLYTE module. Intended Use of the QuikLYTE module : The Na+, K+, Cl and TCO2 methods on the Dimension clinical chemistry system with the QuikLYTE module are in vitro diagnostic tests intended for the quantitative measurement of sodium, potassium, chloride and total carbon dioxide in human serum and heparinized plasma. They are also intended for the quantitative measurement of Na+, K+ and Cl- in urine.

Quantity: 5981

Why Was This Recalled?

A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.

Where Was This Sold?

Worldwide Distribution.