Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IMMULITE/ IMMULITE 1000 Progesterone Recalled by Siemens Healthcare Diagnostics, Inc Due to DHEA-S causes falsely elevated progesterone results.

Date: January 4, 2017
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

IMMULITE/ IMMULITE 1000 Progesterone; Catalog Number: LKPW1; Siemens Material Number: 10381128 Product Usage: For in vitro diagnostic use with the IMMULITE¿2000 Systems Analyzers- for the quantitative measurement of progesterone in serum, as an aid in the diagnosis and treatment of disorders of the ovaries or placenta.

Quantity: 9076

Why Was This Recalled?

DHEA-S causes falsely elevated progesterone results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.