Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with Recalled by Siemens Healthcare Diagnostics, Inc. Due to Potential for Functional Sensitivity to not meet Instructions...

Date: December 7, 2022
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

IMMULITE 2000 Thyroglobulin (US)- For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers (IMMULITE 2000/ IMMULITE 2000 XPi) for the quantitative measurement of thyroglobulin in serum or heparinized plasma, as an aid in monitoring patients who have undergone thyroidectomy. Catalog # L2KTY2(D) Siemens Material Number (SMN): 10381647

Quantity: 717 kits

Why Was This Recalled?

Potential for Functional Sensitivity to not meet Instructions For Use (IFU) claims with the kit lots; control level one may result outside of published ranges; when control results are in range, users may observe increased imprecision with low level patient samples

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report