Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IMMULITE 2000 PSA assay Recalled by Siemens Healthcare Diagnostics, Inc. Due to The IMMULITE 2000 PSA assay for use on...

Date: September 30, 2024
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)

Quantity: 7,332 units

Why Was This Recalled?

The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.

Related Recalls