Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21 Recalled by Siemens Healthcare Diagnostics, Inc Due to The drug fulvestrant (Faslodex¿) may cause falsely elevated...

Date: January 13, 2016
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

IMMULITE 1000 Estradiol Test Code E2 Catalog Number: LKE21, LKE21(D), LKE25 SMN: 10381132, 10702832, 10381142 Product Usage: For in vitro diagnostic use with the IMMULITE¿ and IMMULITE 1000 Analyzers - for the quantitative measurement of estradiol in serum, as an aid in the diagnosis and treatment of various hormonal sexual disorders

Quantity: 9251 units

Why Was This Recalled?

The drug fulvestrant (Faslodex¿) may cause falsely elevated estradiol results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.