Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Dimension(R) IRON Calibrator (DC85) Recalled by Siemens Healthcare Diagnostics, Inc. Due to Siemens Healthcare Diagnostics has received complaints of IRON...

Date: December 20, 2012
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.

Quantity: 5999

Why Was This Recalled?

Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.