Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator Recalled by Siemens Healthcare Diagnostics, Inc. Due to An increase in the number of failed calibration...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.
Affected Products
Dimension Vista System LOCI 8 CALIBRATOR The LOCI 8 Calibrator is an in vitro diagnostic product for the calibration of follicle stimulating hormone (FSH), luteinizing hormone (LH), prolactin (PRL), and estradiol (E2) methods on the Dimension Vista System.
Quantity: 989 (681 USA/308 OUS)
Why Was This Recalled?
An increase in the number of failed calibration events or negative bias with Quality Control (QC) and patient samples when using Dimension Vista LOCI 8 Calibrator lot 7CD076 to calibrate Estradiol (E2) Assay.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Healthcare Diagnostics, Inc.
Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report