Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Dimension Vista¿ MULTI 2 SDIL Recalled by Siemens Healthcare Diagnostics, Inc. Due to The diluent may have an incomplete slit on...

Date: March 21, 2018
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586

Quantity: 1195

Why Was This Recalled?

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Where Was This Sold?

This product was distributed nationwide across all 50 states.