Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Recalled by Siemens Healthcare Diagnostics, Inc. Due to Negative bias with patient samples. The mean patient...

Date: July 18, 2022
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456

Quantity: 129 units US; 159 units OUS

Why Was This Recalled?

Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report