Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Dimension Vista Digoxin (DIGXN) Flex reagent cartridge Recalled by Siemens Healthcare Diagnostics, Inc. Due to The concentrations for Biotin listed in the Non-Interfering...

Date: March 21, 2017
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Dimension Vista Digoxin (DIGXN) Flex reagent cartridge

Quantity: 13,143 units

Why Was This Recalled?

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

Where Was This Sold?

Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa