Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In Recalled by Siemens Healthcare Diagnostics, Inc. Due to Positive bias across the Analytical Measurement Range of...

Date: August 19, 2020
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Dimension LOCI High Sensitivity Troponin I (TNIH) Flex reagent cartridge-IVD-In vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma Catalog Number: RF627, SMN# 10471068

Quantity: 533 US and 1822 OUS

Why Was This Recalled?

Positive bias across the Analytical Measurement Range of the TNIH assay compared to an unaffected control lot. The average bias observed for patient samples using lots EB0255 and EA0178 of the Dimension EXL Troponin compared with a control lot was +25%. A maximum bias of +34% in patient samples around the 99th percentile observed