Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
CentraLink" Data Management System Recalled by Siemens Healthcare Diagnostics, Inc. Due to Siemens Healthcare Diagnostics has determined that the sample...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.
Affected Products
CentraLink" Data Management System; software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Quantity: 1941
Why Was This Recalled?
Siemens Healthcare Diagnostics has determined that the sample query function that includes Instrument or Instrument Group as search criteria may not return all samples from the Dimension Vista systems if onboard aliquot support rules are configured in the CentraLink system.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Healthcare Diagnostics, Inc.
Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report