Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Atellica Sample Handler Connect (SMN 11069018). The Atellica" Solution is Recalled by Siemens Healthcare Diagnostics, Inc. Due to Atellica Solution system software (SW) version 1.13 and...

Date: June 29, 2018
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Atellica Sample Handler Connect (SMN 11069018). The Atellica" Solution is a multi-component system for in vitro diagnostic testing of clinical specimens.

Quantity: 35

Why Was This Recalled?

Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks¿ and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.