Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
Atellica IM anti-CCP IgG (aCCP) Assay Siemens Material Number: 10732998 Recalled by Siemens Healthcare Diagnostics, Inc Due to Atellica IM anti-CCP IgG (aCCP) Test Definition (TDef)...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.
Affected Products
Atellica IM anti-CCP IgG (aCCP) Assay Siemens Material Number: 10732998 (100 Test) Test Version1.0 - Product Usage: The ADVIA Centaur¿ Anti-CCP IgG (aCCP) assay is for in vitro diagnostic use in the semi-quantitative determination of the IgG class of autoantibodies specific to cyclic citrullinated peptide (CCP) in human serum and plasma (K2-EDTA, lithium heparin) using the ADVIA Centaur XP system. Detection of anti-CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis (RA) and should be used in conjunction with other clinical information. Autoantibody levels represent one parameter in a multi-criteria diagnostic process, encompassing both clinical and laboratory-based assessments. [Note: the Atellica IM anti-CCP IgG (aCCP) assay was introduced for sale through the Replacement Reagent Policy.]
Quantity: 169 kits
Why Was This Recalled?
Atellica IM anti-CCP IgG (aCCP) Test Definition (TDef) Mitigation Failure-reagent carryover mitigation for the Testosterone II (TSTII) assay is unable to be completed as defined in the aCCP Test Definition (TDef). The issue occurs on an Atellica IM Analyzer and is isolated to TDef Version 1.0. FOR OUS: Atellica IM anti-CCP IgG (aCCP) and Pregnancy-Associated Plasma Protein-A (PAPP-A) Test Definition (TDef) Mitigation Failure
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Healthcare Diagnostics, Inc
Siemens Healthcare Diagnostics, Inc has 118 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report