Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

(1) Atellica Data Manager software version 1.0 Siemens Material Number Recalled by Siemens Healthcare Diagnostics, Inc. Due to Unexpected interface driver behavior identified (QC) Results May...

Date: May 15, 2020
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

(1) Atellica Data Manager software version 1.0 Siemens Material Number (SMN): 11314237 (2) Atellica Data Manager software version 1.1 Siemens Material Number (SMN): 11316888 Product Usage: are clinical laboratory information modules which consolidate test orders and test results, provide test ordering and priority routing information to automation, apply rules to test orders and auto-verification rules for test results, and provide integrated management of quality control of test system.

Quantity: 70 units

Why Was This Recalled?

Unexpected interface driver behavior identified (QC) Results May Be Assigned to an Incorrect Control Lot Number, and lead to the reporting of erroneous patient results if the QC failed but appeared to be passing and the issue is not detected during QC review

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.