Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Atellica CI Analyzer. Catalog Numbers: 10947347. Recalled by Siemens Healthcare Diagnostics, Inc. Due to Potential that the IMT Diluent volume remaining (%...

Date: November 4, 2024
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

Atellica CI Analyzer. Catalog Numbers: 10947347.

Quantity: 559 units

Why Was This Recalled?

Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

Where Was This Sold?

This product was distributed nationwide across all 50 states.