Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

N Antiserum to Human IgG Recalled by SIEMENS HEALTHCARE DIAGNOSTICS INC Due to N Antiserum to Human IgG are not meeting...

Date: August 30, 2022
Company: SIEMENS HEALTHCARE DIAGNOSTICS INC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact SIEMENS HEALTHCARE DIAGNOSTICS INC directly.

Affected Products

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Quantity: 10,201 packs

Why Was This Recalled?

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Where Was This Sold?

This product was distributed nationwide across all 50 states.