Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.
ADVIA Chemistry XPT Systems Recalled by Siemens Healthcare Diagnostics, Inc. Due to Multiple Software issues. Auto Start-UP Fail, Calibration Interval...
What You Should Do
Stop using this product immediately. Do not consume, use, or distribute it.
Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.
Affected Products
ADVIA Chemistry XPT Systems, Software Version 1.0.3, SMN 11127538 Clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.
Quantity: 87 units
Why Was This Recalled?
Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS) communication/Laboratory Automation (LAS) issue, Printer Driver Resets, ISE Calibration Ranges are too conservative for Urine Sodium, Archiving and deletion may fail and Workstation services may restart.
Where Was This Sold?
This product was distributed nationwide across all 50 states.
About Siemens Healthcare Diagnostics, Inc.
Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.
Related Recalls
EndoVive Safety PEG Kit Recalled by Boston Scientific Corporation Due to Recall of kits which utilized ChloraPrep Triple Swabsticks...
Boston Scientific Corporation · February 10, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Artelon FlexBand Dynamic Matrix Ref: 31057 Recalled by International Life Sciences Due to Augmentation devices failed bacterial endotoxin testing.
International Life Sciences · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Owner's Booklets and Instructions for Use that are used with Recalled by Trividia Health, Inc. Due to The system labeling (user manual and online labeling)...
Trividia Health, Inc. · February 6, 2026
Data sourced from the FDA (Device). Last updated March 26, 2026. View original report