Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use Recalled by Siemens Healthcare Diagnostics, Inc. Due to Urea Nitrogen reagent kit demonstrated an increased incidence...

Date: March 27, 2018
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems

Quantity: N/A

Why Was This Recalled?

Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.

Where Was This Sold?

This product was distributed nationwide across all 50 states.