Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ADVIA Chemistry ToxAmmonia Calibrator- In vitro diagnostic use in the Recalled by Siemens Healthcare Diagnostics, Inc. Due to Reassignment of the Atellica CH Toxicology Calibrators (TOX...

Date: April 5, 2023
Company: Siemens Healthcare Diagnostics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

ADVIA Chemistry ToxAmmonia Calibrator- In vitro diagnostic use in the calibration of ethanol (ETOH_2), ammonia (AMM), acetaminophen (ACET), and salicylate (SAL) assays on the ADVIA¿ Chemistry systems Siemens Material Number (SMN): 10309217

Quantity: 447 unit US; 1537 units OUS

Why Was This Recalled?

Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias

Where Was This Sold?

This product was distributed nationwide across all 50 states.