Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ADVIA Chemistry Hemoglobin A1c_3 Recalled by Siemens Healthcare Diagnostics, Inc. Due to Reagent in these lots may demonstrate an increased...

Date: September 24, 2015
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

ADVIA Chemistry Hemoglobin A1c_3, A1c_3M Reagent kits. Used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems. For in vitro diagnostic use in the quantitative determination of Hemoglobin A1c, a diabetes marker, in whole blood on the ADVIA Chemistry systems.

Quantity: 3031

Why Was This Recalled?

Reagent in these lots may demonstrate an increased occurrence of high %HbA1c bias, a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The maximum bias was observed at higher %HbA1c concentrations. QC samples may exhibit a similar bias.

Where Was This Sold?

This product was distributed to 10 states: AZ, CA, GA, IN, LA, MI, MN, NM, NY, OH

Affected (10 states)Not affected