Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

ADVIA Chemistry Enzyme 2 Calibrator Recalled by Siemens Healthcare Diagnostics, Inc. Due to Siemens Healthcare Diagnostics has confirmed a positive bias...

Date: February 23, 2017
Company: Siemens Healthcare Diagnostics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc. directly.

Affected Products

ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Quantity: 314 calibrators

Why Was This Recalled?

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc.

Siemens Healthcare Diagnostics, Inc. has 452 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report