Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS Recalled by Siemens Healthcare Diagnostics, Inc Due to Samples containing fluorescein may show interference with the...

Date: April 10, 2014
Company: Siemens Healthcare Diagnostics, Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Siemens Healthcare Diagnostics, Inc directly.

Affected Products

ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 10378883.

Quantity: 16,302 kits

Why Was This Recalled?

Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Siemens Healthcare Diagnostics, Inc

Siemens Healthcare Diagnostics, Inc has 118 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report