Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17 Recalled by Shimadzu Medical Systems Due to It was found that the irradiated x-ray may...

Date: January 10, 2022
Company: Shimadzu Medical Systems
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Shimadzu Medical Systems directly.

Affected Products

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

Quantity: 12 systems

Why Was This Recalled?

It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Where Was This Sold?

This product was distributed to 37 states: AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IA, KS, KY, LA, MA, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, OH, OR, PA, RI, SC, SD, TN, TX, WA, WV

Affected (37 states)Not affected

About Shimadzu Medical Systems

Shimadzu Medical Systems has 25 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report